Tygon® Medical Tubing - S-95-E |

Typical Applications:
-
Extracorporeal
membrane oxygenation (ECMO)
pumps
-
Kidney dialysis
pumps
-
Peristaltic pumps
-
Blood and IV
solutions
|
TYGON®
S-95-E is a medical grade, clear, non-DEHP polymer
tubing offering superior peristaltic pump life and
exceptional flexibility. TYGON
S-95-E is
specifically designed for Extracorporeal Membrane
Oxygenation (ECMO) and other critical care
procedures, including kidney dialysis. With its
longer pump life, TYGON
S-95-E saves time and money
by reducing the number of tubing changeovers needed.
TYGON
S-95-E features extremely low extractables and
is ideal for use in all peristaltic pumping
applications involving blood transfer. The standard
of the industry for life-saving ECMO procedures,
TYGON
S-95-E tubing provides high quality and
consistent performance that's unmatched by any other
product on the market today.
Biocompatibility
TYGON® S-95-E is made
from a biocompatible polymer material developed
specifically for medical applications and meets USP
Class VI standards. It meets ISO10993 and European
Pharmacopoeia 3.1.1.2 standards. |
|
|
|
Features and Benefits:
-
Superior peristaltic pump
life
-
Non-DEHP formulation
-
Ideal for contact with
blood
-
Meets USP Class VI,
European Pharmacopoeia 3.1.1.2 and FDA criteria
-
Meets ISO 10993 standards
-
Withstands both EtO and
gamma sterilization
-
Registered with the FDA
under Master File MAF106
|
|
|
TYGON® S-95-E
Typical Physical Properties*
|
|
Property |
ASTM
Method |
Value
or Rating |
|
Durometer Hardness (Shore A), 15S |
D2240-02 |
66 |
|
Tensile Strength, psi (MPa) (at break) |
D412-98 |
2,000
(13.8) |
|
Ultimate Elongation, % |
D412-98 |
390 |
| Tear
Resistance, lb-f/inch (kN/m) |
D624-91,
Die C |
130
(22.7) |
|
Scecific Gravity |
D792-02 |
1.16 |
|
Color |
n/a |
Clear |
|
|
*Unless
otherwise noted, all tests were conducted at room
temperature (73˚F). Values shown were determined on 0.075"
thick extruded strip or 0.075" thick molded ASTM plaques or
molded ASTM durometer buttons. |
|
|
|
TYGON® S-95-E
Product Characteristics |
TYGON® S-95-E Sterilization Methods |
| |
|
|
ISO 10993 - Cytotoxicity |
Pass |
|
ISO 10993 - Hemocompatability |
Pass |
|
ISO 10993 - Pyrogens |
Pass |
|
ISO 10993 - Sensitization |
Pass |
|
ISO 10993 - Irritation |
Pass |
|
ISO 10993 - In Vitro Hemolysis |
Pass |
|
ISO 10993 - Genotoxicity |
Pass |
|
USP Fungistasis |
Pass |
USP Physicochemical Testing for
Elastomeric Closures |
Pass |
|
USP Class VI |
Pass |
|
European Pharmacopoeia 3.1.1.2 |
Pass |
|
C3a Complement Activation Assay |
Pass |
|
|
Autoclavable |
Yes |
|
Gas |
Yes |
|
Radiation |
Yes (2.5 MRad) |
|
| |
| |
TYGON® S-95-E Relative Chemical
Resistance Properties* |
| |
|
Acids
|
Bases |
|
|
concentrated |
F |
|
concentrated |
E |
|
Salts |
E |
|
medium |
E |
medium |
E |
Alcohols |
F |
|
weak |
E |
weak |
E |
Ketones |
U |
|
E=Excellent, F=Fair, U=Unsatisfactory
*All tests conducted at room